Thin Layer Chromatography - TLC

TLC is a simple, quick, and inexpensive procedure that gives the chemist a quick answer as to how many components are in a mixture. TLC is also used to support the identity of a compound in a mixture when the Rf of a compound is compared with the Rf of a known compound (preferrably both run on the same TLC plate).

A TLC plate is a sheet of glass, metal, or plastic which is coated with a thin layer of a solid adsorbent (usually silica or alumina). A small amount of the mixture to be analyzed is spotted near the bottom of this plate. The TLC plate is then placed in a shallow pool of a solvent in a developing chamber so that only the very bottom of the plate is in the liquid. This liquid, or the eluent, is the mobile phase, and it slowly rises up the TLC plate by capillary action.

As the solvent moves past the spot that was applied, an equilibrium is established for each component of the mixture between the molecules of that component which are adsorbed on the solid and the molecules which are in solution. In principle, the components will differ in solubility and in the strength of their adsorption to the adsorbent and some components will be carried farther up the plate than others. When the solvent has reached the top of the plate, the plate is removed from the developing chamber, dried, and the separated components of the mixture are visualized. If the compounds are colored, visualization is straightforward. Usually the compounds are not colored, so a UV lamp is used to visualize the plates. (The plate itself contains a fluor which fluoresces everywhere except where an organic compound is on the plate.)

The procedure for TLC, explained in words in the above paragraphs, is illustrated with photographs on the TLC Procedure page.

The Indiana University site has an animated graphics movie on TLC procedure.

TLC Adsorbent

In the teaching labs at CU Boulder, we use silica gel plates (SiO2) almost exclusively. (Alumina (Al2O3) can also be used as a TLC adsorbent.) The plates are aluminum-backed and you can cut them to size with scissors. Our plates are purchased ready-made from EM Sciences or from Scientific Adsorbents. The adsorbent is impregnated with a fluor, zinc sulfide. The fluor enables most organic compounds to be visualized when the plate is held under a UV lamp. In some circumstances, other visualization methods are used, such as charring or staining.

TLC Solvents or Solvent Systems

Choosing a solvent is covered on the Chromatography Overview page. The charts at the bottom of that page are particularly useful.

Interactions of the Compound and the Adsorbent

The strength with which an organic compound binds to an adsorbent depends on the strength of the following types of interactions: ion-dipole, dipole-dipole, hydrogen bonding, dipole induced dipole, and van der Waals forces. With silica gel, the dominant interactive forces between the adsorbent and the materials to be separated are of the dipole-dipole type. Highly polar molecules interact fairly strongly with the polar Si—O bonds of these adsorbents and will tend to stick or adsorb onto the fine particles of the adsorbent while weakly polar molecules are held less tightly. Weakly polar molecules thus generally tend to move through the adsorbent more rapidly than the polar species. Roughly, the compounds follow the elution order given on the Chromatography Overview page.

The Rf value

Rf is the retention factor, or how far up a plate the compound travels. See the Rf page for more details:

Visualizing the Spots

If the compounds are colored, they are easy to see with the naked eye. If not, a UV lamp is used (see the Procedure page).

Troubleshooting TLC

All of the above (including the procedure page) might sound like TLC is quite an easy procedure. But what about the first time you run a TLC, and see spots everywhere and blurred, streaked spots? As with any technique, with practice you get better. One thing you have to be careful Examples of common problems encountered in TLC: